Biomechanical analysis of intact versus ruptured Poly Implant Prothèse breast implants

Author:

Ramião N. G.1ORCID,Martins P. S.1,Barroso M. L.2,Santos D. C.2,Fernandes A. A.1

Affiliation:

1. INEGI, LAETA, Faculty of Engineering of the University of Porto, Campus da FEUP, Rua Dr. Roberto Frias, 400, 4200-465 Porto, Portugal

2. Department of Plastic Surgery, Gaia Hospital Center, Vila Nova de Gaia, Portugal

Abstract

Despite there being many studies that have evaluated breast implant rupture, there is no consensus about causes and incidence. Most studies lack a multifactorial analysis of what causes breast implants to rupture. To fill this gap, an experimental protocol was developed to compare ruptured and intact Poly Implant Prothèse (PIP) breast implants from the same woman. These conditions guarantee that the physical/biological variables are the same for each pair of ruptured and intact implants. A total of 1008 samples from 22 PIP explants (11 intact and 11 ruptured) and three control PIP implants were analysed. The mechanical properties (tensile strength) of the ruptured and intact implants were compared according to brand, lot, implantation time and demographic conditions. In general, statistically significant differences were found between the intact and ruptured PIP implants. Ruptured implants were thinner (0.73 ± 0.10 mm versus 0.91 ± 0.11 mm) and weaker (7.42 ± 2.65 MPa versus 9.59 ± 2.37 MPa) than intact implants. Intact and ruptured implants have shown distinct mechanical behaviours and variations in thickness. Our understanding is that these differences may be associated with the typical manufacturing process of breast implant shells. These results stress the importance of thorough control of the shell thickness. Given its relevance, shell thickness should be used as a quality control measure for homologation purposes. Thus, the homogeneity of the shell should be considered as a relevant parameter during the manufacturing process. This will translate into an improved quality of life for patients and will potentiate safer and longer lasting products.

Funder

Ministério da Ciência, Inovação e do Ensino Superior, FCT - FundaçãoparaaCiência e a Tecnologia, Portugal,

FundoEuropeu deDesenvolvimentoRegional (FEDER) andbyNational Funds

SciTech - Science and Technology for Competitive and Sustainable Industries

MImBI

Publisher

The Royal Society

Subject

Biomedical Engineering,Biomaterials,Biochemistry,Bioengineering,Biophysics,Biotechnology

Reference34 articles.

1. Rupture and Aging of Silicone Gel Breast Implants

2. Strength of Silicone Breast Implants

3. Mechanical Analysis of Explanted Silicone Breast Implants

4. Surface and mechanical analysis of explanted Poly Implant Prosthèse silicone breast implants

5. On the failure of silicone breast implants: new insights by mapping the mechanical properties of implant shells;Schubert DW;Soc. Chem. Industry,2013

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Aging of non-implanted Natrelle™ gel breast implants;Journal of the Mechanical Behavior of Biomedical Materials;2023-11

2. Bioengineering in women's health: part I;Interface Focus;2019-06-14