Abstract
A radial heat-flow apparatus has been constructed in Pyrex glass and used for liquid thermal conductivity investigations. The liquid under test fills the 1.25mm space between concentric cylinders and a measured quantity of heat flows inwards. After calibration using liquids of known thermal conductivity, new values are obtained for pyridine and three chlorofluorocarbon oils. The apparatus can also be used for direct determinations when allowance is made for the temperature drop occurring in each of the glass walls. For this purpose, a determination of the thermal conductivity of the glass was made using a similar method. A discussion of the results is included.
Reference2 articles.
1. Birch F . & Clark H . 1940 Amer. J .
2. Bullard E. C. & N iblett B. R . 1951 Sci.238 529.
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献