Author:
Wang Shanshan,Zeng Liang,Song Junjie,Zhou Liang,Ding Mingyue,Yuchi Ming
Abstract
Ultrasound computed tomography (USCT) has important clinical application prospect in breast cancer screening and early diagnosis. In this paper, six kinds of coherence factor-like beamforming methods have been applied to improve the image quality for USCT, including coherence factor
(CF), phase coherence factor (PCF), sign coherence factor (SCF), phasor dispersion based coherence factor (PDCF), spatial smoothed coherence factor (SSCF) and spatio-temporally smoothed coherence factor (STSCF). The mentioned methods were verified with the radio-frequency (RF) data of the
breast phantom captured by the USCT system developed in the Medical Ultrasound Laboratory. The ring-type transducer has 1024 elements with a center frequency of 2.5 MHz. Experimental results show that the reconstructed images of the breast phantom by the CF gets the highest contrast to noise
ratio (CNR), but overall image brightness reduces significantly. PCF gets the lowest variance and provides a more homogenous background. STSCF beamforming method can improve the robustness of the PCF and having the ability to suppress clutter while significant removal of black region artifacts.
For practical application, these coherence factor-like beamforming methods can be implemented with low computational complexity.
Publisher
American Scientific Publishers
Subject
Health Informatics,Radiology Nuclear Medicine and imaging