Abstract
Aiming at exploring biomedical signal acquisition equipment used in human motion heart rate monitoring, the research on the related hardware design and signal processing method was carried out. A biomedical signal acquisition device based on photoplethysmography (PPG) is designed, and
the equipment was applied to acquire PPG signals and acceleration sensor signals under different motion states. The analysis of the experimental data showed that, the fusion method of the acceleration sensing information in the motion artifact removal method is perfected. The effectiveness
of the baseline drift removal algorithm, motion artifact removal algorithm and dynamic heart rate monitoring algorithm was verified by reconstructing the signal quality evaluation index. To sum up, taking MINDRAY VS-800 as a reference device, it is compared with the adaptive filtering technology
in terms of signal quality, BPM detection results and algorithm complexity, and better results are finally obtained.
Publisher
American Scientific Publishers
Subject
Health Informatics,Radiology, Nuclear Medicine and imaging
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献