Аnalysis of the technical and dosimetric characteristics of the brachytherapy afterloaders. Ring applicator

Author:

Kazlouski D. I.1,Titovich Y. V.2,Kazlouskaya Y. I.3

Affiliation:

1. International Sakharov Environmental Institute of Belorussian State University

2. ADANI

3. N.N. Alexandrov National Cancer Centre

Abstract

A study of the technical and dosimetry characteristics of brachytherapy afterloaders and applicators was carried out. Ring applicator has been taken as an example, the correctness of positioning of a radiation source (RS) inside the applicators was tested as part of the comissioning and quality control procedure of the brachytherapy applicators. The magnitudes of inconsistencies in the position of RS were established when planning and implementing treatment plans for radiation therapy. The identification of the values of the discrepancy was carried out using the obtained X-ray images of the applicator at the time of the implementation of the irradiation plans. The treatment plan was a sequential positioning of RS in the body of the applicator in each active position with a minimum step from the tip to the vaginal part of the applicator. The X-ray image was obtained by locating the source sequentially at each active position of the applicator. When carrying out dosimetric planning, 3 methods of applicator reconstruction were used. The analysis revealed that the applicator reconstruction method affects the magnitude of the discrepancy in determining the position of the source in the lumen of the applicator ring. Using the methods of statistical analysis, the mean, median, maximum and minimum values of the detected deviations were calculated. The results are presented in the form of tables and graphs for all investigated stop positions of IRS. Based on the results of the study, we consider it expedient to carry out quality control procedures when putting applicators into clinical operation. Based on the results obtained, we consider it acceptable to conduct a quality control procedure for the positioning accuracy of radiation sources in the applicators at least 1 time per month. Taking into account the results of the study when carrying out dosimetric planning will improve the quality of the irradiation sessions using the brachytherapy method, thereby improving the quality of oncological care for the population.

Publisher

Belarusian State University of Informatics and Radioelectronics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3