Abstract
Abstract
It was previously shown that the experimental measurement of the lifetime of gas-phase negative ions formed in autoionization states during resonant electron capture by molecules is inherently fraught with inaccuracy due to adsorption of ions on the walls of the ionization chamber, with some of them surviving there and staying stabilized as ‘eternal’ till they eventually desorb back into the vacuum and merge with the total ion flow; this leads to an upward distortion of lifetime measurements. This paper describes a technique for removing desorbed ions from the total flow using an ion filter that lets only gas-phase ions pass into the detection system but detains the desorbed ions due to the differences among their kinetic energies.
Publisher
Uspekhi Fizicheskikh Nauk (UFN) Journal
Subject
General Physics and Astronomy
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献