Accuracy of blood-pressure monitors owned by patients with hypertension (ACCU-RATE study): a cross-sectional, observational study in central England

Author:

Hodgkinson James A,Lee Mei-Man,Milner Siobhan,Bradburn Peter,Stevens Richard,Hobbs FD Richard,Koshiaris Constantinos,Grant Sabrina,Mant Jonathan,McManus Richard J

Abstract

BackgroundHome blood-pressure (BP) monitoring is recommended in guidelines and is increasingly popular with patients and health professionals, but the accuracy of patients’ own monitors in real-world use is not known.AimTo assess the accuracy of home BP monitors used by people with hypertension, and to investigate factors affecting accuracy.Design and settingCross-sectional, observational study in urban and suburban settings in central England.MethodPatients (n = 6891) on the hypertension register at seven practices in the West Midlands, England, were surveyed to ascertain whether they owned a BP monitor and wanted it tested. Monitor accuracy was compared with a calibrated reference device at 50 mmHg intervals between 0–280/300 mmHg (static pressure test); a difference from the reference monitor of +/−3 mmHg at any interval was considered a failure. Cuff performance was also assessed. Results were analysed by frequency of use, length of time in service, make and model, monitor validation status, purchase price, and any previous testing.ResultsIn total, 251 (76%, 95% confidence interval [95% CI] = 71 to 80%) of 331 tested devices passed all tests (monitors and cuffs), and 86% (CI] = 82 to 90%) passed the static pressure test; deficiencies were, primarily, because of monitors overestimating BP. A total of 40% of testable monitors were not validated. The pass rate on the static pressure test was greater in validated monitors (96%, 95% CI = 94 to 98%) versus unvalidated monitors (64%, 95% CI = 58 to 69%), those retailing for >£10 (90%, 95% CI = 86 to 94%), those retailing for ≤£10 (66%, 95% CI = 51 to 80%), those in use for ≤4 years (95%, 95% CI = 91 to 98%), and those in use for >4 years (74%, 95% CI = 67 to 82%). All in all, 12% of cuffs failed.ConclusionPatients’ own BP monitor failure rate was similar to that demonstrated in studies performed in professional settings, although cuff failure was more frequent. Clinicians can be confident of the accuracy of patients’ own BP monitors if the devices are validated and ≤4 years old.

Publisher

Royal College of General Practitioners

Subject

Family Practice

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3