Affiliation:
1. University of Sao Paulo, Brazil
2. Universidade Federal do Triangulo Mineiro, Brazil
Abstract
OBJECTIVE: To evaluate the performance of manual resuscitators (MRs) used in Brazil in accordance with international standards. METHODS: Using a respiratory system simulator, four volunteer physiotherapists employed eight MRs (five produced in Brazil and three produced abroad), which were tested for inspiratory and expiratory resistance of the patient valve; functioning of the pressure-limiting valve; and tidal volume (VT) generated when the one-handed and two-handed techniques were used. The tests were performed and analyzed in accordance with the American Society for Testing and Materials (ASTM) F920-93 criteria. RESULTS: Expiratory resistance was greater than 6 cmH2O . L−1 . s−1 in only one MR. The pressure-limiting valve, a feature of five of the MRs, opened at low pressures (< 17 cmH2O), and the maximal pressure was 32.0-55.9 cmH2O. Mean VT varied greatly among the MRs tested. The mean VT values generated with the one-handed technique were lower than the 600 mL recommended by the ASTM. In the situations studied, mean VT was generally lower from the Brazilian-made MRs that had a pressure-limiting valve. CONCLUSIONS: The resistances imposed by the patient valve met the ASTM criteria in all but one of the MRs tested. The pressure-limiting valves of the Brazilian-made MRs usually opened at low pressures, providing lower VT values in the situations studied, especially when the one-handed technique was used, suggesting that both hands should be used and that the pressure-limiting valve should be closed whenever possible.
Subject
Pulmonary and Respiratory Medicine
Reference20 articles.
1. Equipment Theory for Respiratory Care;White GC,2005
2. Fatores que afetam a ventilação com o reanimador manual autoinflável: uma revisão sistemática;Oliveira PM;Rev Paul Pediatr,2011
3. The effects of three manual hyperinflation techniques on pattern of ventilation in a test lung model;Maxwell L;Anaesth Intensive Care,2002
4. Physiotherapy in intensive care: towards an evidence-based practice;Stiller K;Chest,2000
5. Oxygen outflow delivered by manually operated self-inflating resuscitation bags in patients breathing spontaneously;Godoy AC;J Bras Pneumol,2008
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献