Abstract
Three certified samples of different matrices (Soil-5, SL-1/IAEA and SARM-4/SABS) were quantitatively analysed by energy dispersive X-ray fluorescence with radioisotopic excitation. The observed errors were about 10-20% for the majority of the elements and less than 10% for Fe and Zn in the Soil-5, Mn in SL-1, and Ti, Fe and Zn in SARM-4 samples. Annular radioactive sources of Fe-55 and Cd-109 were utilized for the excitation of elements while a Si(Li) semiconductor detector coupled to a multichannel emulation card inserted in a microcomputer was used for the detection of the characteristic X-rays. The fundamental parameters method was used for the determination of elemental sensitivities and the irradiator or transmission method for the correction of the absorption effect of characteristic X-rays of elements on the range of atomic number 22 to 42 (Ti to Mo) and excitation with Cd-109. For elements in the range of atomic number 13 to 23 (Al to V) the irradiator method cannot be applied since samples are not transparent for the incident and emergent X-rays. In order to perform the absorption correction for this range of atomic number excited with Fe-55 source, another method was developed based on the experimental value of the absorption coefficients, associated with absorption edges of the elements.
Subject
Agronomy and Crop Science,Animal Science and Zoology
Cited by
15 articles.
订阅此论文施引文献