Indirect calorimetry can be used to measure cardiac output in septic patients?

Author:

Martins Maria Auxiliadora1,Coletto Francisco Antônio1,Campos Antônio Dorival2,Basile-Filho Anibal1

Affiliation:

1. University of São Paulo, Brasil

2. University of São Paulo, Brazil

Abstract

PURPOSE: The aim of this study was to compare two different cardiac output (CO) monitoring systems based on the thermodilution principle (Thermo-CO) and indirect calorimetry (Fick mixed-CO) in septic patients. METHODS: Prospective study in septic patients admitted in an intensive care unit of a university hospital. Nineteen patients aged on average 45.4 ± 21.5 years were enrolled in the study. Four series of hourly measurements by the two techniques were carried out simultaneously. RESULTS: No significant differences were observed between Thermo-CO and Fick mixed-CO (7.0 ± 1.8 L.min-1 and 6.4 ± 1.7 L.min-1.). Parallel analysis of Fick mixed-CO and Fick atrial-CO was performed introducing a correction factor for the eight atrial samples in order to adjust the values of oxygen saturation obtained from atrial blood (Fick corrected atrial-CO) to those obtained from mixed venous blood. No significant differences could be detected between Fick mixed-CO and Fick corrected atrial-CO. The correlation coefficients of Thermo CO/Fick mixed-CO and Fick mixed-CO/Fick corrected atrial-CO were 0.84 and 0.94, respectively. CONCLUSION: We observed that the agreement between the two methods was satisfactory on the basis of the decisions made for treatment. Indirect calorimetry is useful to measure CO in patients with septic shock.

Publisher

FapUNIFESP (SciELO)

Subject

Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3