Endoscope reprocessing using glutaraldehyde in endoscopy services of Goiânia, Brazil

Author:

Barbosa Jackeline Maciel1,Souza Adenicía Custódia Silva2,Tipple Anaclara Ferreira Veiga2,Pimenta Fabiana Cristina2,Leão Lara Stefania Netto de Oliveira2,Silva Silvia Rita Marin Caninni3

Affiliation:

1. Municipal Secretary of Goiânia

2. UFG

3. Ribeirão Preto São Paulo University, Brazil

Abstract

CONTEXT: The endoscopic procedure safety depends on the use of an adequately reprocessed device which quality is related to each of its operational steps. OBJECTIVE: To characterize the reprocessing of endoscopes using glutaraldehyde in endoscopy services METHODS: Study was conducted by observing the reprocessing of 60 endoscopes from 20 medical practices of the municipality of Goiânia, GO, central area of Brazil. RESULTS: This study showed failure in all reprocessing steps. The pre-washing was performed in 24 (40.0%) of the endoscope. In the cleaning steps, was identify the improper use of enzymatic detergent, and in 27 (45.0%) cases, the brushing of internal channels was not performed. All 60 endoscopes were submitted to this disinfectant. However, for 33 (55.0%) of the cases the internal channels was not filled. The total immersion of endoscope in the glutaraldehyde was not performed in 39 (65.0%) cases. The recommended minimum total immersion time for exposure to 2% glutaraldehyde solution was followed only for 12 (20.0%) endoscopes. There was no filter for water treatment used in the rinse of most endoscopes 54 (90.0%) and to dry the internal channels only 6 (10.0%) of them used compressed air. Adequate storing conditions were identified. CONCLUSION: Considering the particularities of the endoscope and its reprocessing, it is imperative to establish protocols to ensure the quality of the disinfection and the prevention of cross-contamination.

Publisher

FapUNIFESP (SciELO)

Subject

Gastroenterology

Reference25 articles.

1. Association for Professionals in Infection Control and Epidemiology (APIC): Guideline for infection prevention and control in flexible endoscopy;Alvarado CJ;Am J Infect Control,2000

2. A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure;Bisset L;Am J Infect Control,2006

3. Portaria nº 593 de 25 de agosto de 2000: Aprova o Regulamento Técnico para Serviço de Endoscopia Digestiva e Respiratória,2000

4. Informe Técnico nº 04/07 de 2007: Glutaraldeído em estabelecimentos de assistência a saúde fundamentos para a utilização,2007

Cited by 12 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献