Infection of expanded polytetrafluoroethylene and Dacron-coated stents with Staphylococcus epidermidis: an experimental study in pigs

Author:

Dutra Clandio de Freitas1ORCID,Pereira Adamastor Humberto2,Wollheim Claudia1,Pongiluppi Rodrigo1,Fellini Roberto1,Gomes Junior Sérgio Ventura1,Nonemacher Henrique3

Affiliation:

1. Universidade de Caxias do Sul, Brasil

2. Universidade Federal do Rio Grande do Sul, Brasil

3. Hospital Santa Mônica Erechim, Brasil

Abstract

Abstract Background Diagnosis of the etiologic agent of endoprosthesis infections is essential to enable treatment, since these infections constitute important complications of endovascular procedures. Sonication of explanted tissue and materials is a technique that can be used to facilitate detection of biofilm-producing bacteria. Objectives To evaluate infection of pigs' aortas after implantation of nitinol stents coated with polytetrafluoroethylene (ePTFE) or Dacron, previously infected with biofilm-producing Staphylococcus epidermidis. Intimal thickening and the inflammatory response in the aortic wall were also evaluated. Methods 11 ePTFE-coated nitinol stents and 10 Dacron stents infected with S. epidermidis strains were implanted in the infrarenal aorta of 21 8-week-old pigs. After 2 weeks, the aorta containing the stents was removed. A vortex mixer and ultrasound were used to homogenize the samples and remove the biofilm. Subsequently, the number of colony-forming units was counted. Results There were no significant differences between the two groups in terms of the number of colony-forming units or of inflammation in the arterial wall. With the exception of one specimen from the Dacron group, all aortic stent cultures were positive for S. epidermidis. Conclusions There were no significant differences in the inflammatory response or infection rate between ePTFE and Dacron-coated stents actively infected with biofilm-producing S. epidermidis. Intimal thickening and the inflammatory response to infection of endoprostheses were similar. These results suggest that the two most widely used stent lining materials have a similar infection rate.

Publisher

FapUNIFESP (SciELO)

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3