A new spectrophotometric method to detect residual amounts of peroxide after reprocessing hemodialysis filters

Author:

Oliveira Moacir de1,Dalboni Maria Aparecida1,Iizuka Ilson Jorge1,Manfredi Silvia Regina1,Guimarães Nadia Karina1,Andreoli Maria Claudia Cruz1,Matos Ana Cristina Carvalho1,Batista Marcelo Costa1,Santos Bento Fortunato Cardoso1,Cendoroglo Neto Miguel1

Affiliation:

1. Universidade Federal de São Paulo – UNIFESP, Brazil

Abstract

ABSTRACT Objective: Reuse of hemodialysis filters is a standard practice and the sterilizing chemical most often employed is peracetic acid. Before starting the dialysis session, filters and lines are checked for residual levels of peracetic acid by means of a non-quantitative colorimetric test that is visually interpreted. The objective of this study was to investigate a new quantitative spectrophotometric test for detection of peracetic acid residues. Methods: Peracetic acid solutions were prepared in concentrations ranging from 0.01 to 10 ppm. A reagent (potassium-titanium oxide + sulfuric acid) was added to each sample in proportions varying from 0.08 to 2.00 drops/mL of solution. Optical densities were determined in a spectrophotometer using a 405-nm filter and subjected to visual qualitative test by different observers. Results: A relation between peroxide concentrations and respective optical densities was observed and it was linear with R2 > 0.90 for all reagent/substrate proportions. The peak optical densities were obtained with the reagent/substrate ratio of 0.33 drops/mL, which was later standardized for all further experiments. Both qualitative and quantitative tests yielded a specificity of 100%. The quantitative test was more sensitive than the qualitative test and resulted in higher positive and negative predictive values. There was a difference between observers in the qualitative test and some samples with significant amounts of peroxide were not detected. Conclusion: A quantitative spectrophotometric test may improve detection of residues of peracetic acid when compared to the standard visual qualitative test. This innovation may contribute to the development of safer standards for reuse of hemodialysis filters.

Publisher

FapUNIFESP (SciELO)

Subject

General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3