Effects of aluminum-copper alloy filtration on photon spectra, air kerma rate and image contrast

Author:

Gonçalves Andréa1,Rollo João Manuel Domingos de Almeida2,Gonçalves Marcelo1,Haiter Neto Francisco3,Bóscolo Frab Norberto3

Affiliation:

1. UNESP, Brazil

2. USP, Brazil

3. UNICAMP, Brazil

Abstract

This study evaluated the performance of aluminum-copper alloy filtration, without the original aluminum filter, for dental radiography in terms of x-ray energy spectrum, air kerma rate and image quality. Comparisons of various thicknesses of aluminum-copper alloy in three different percentages were made with aluminum filtration. Tests were conducted on an intra-oral dental x-ray machine and were made on mandible phantom and on step-wedge. Depending on the thickness of aluminum-copper alloy filtration, the beam could be hardened and filtrated. The use of the aluminum-copper alloy filter resulted in reductions in air kerma rate from 8.40% to 47.33%, and indicated the same image contrast when compared to aluminum filtration. Aluminum-copper alloy filtration may be considered a good alternative to aluminum filtration.