Abstract

ABSTRACT Objective: To investigate infusion pumps start-up delay according to different brands of infusion pumps, flow rates and intravenous sets priming techniques. Method: The experimental study simulated clinical practice under controlled conditions, using a 50 mL syringe with NaCl 0.9% solution, two syringe infusion pumps (A and B), six rates (0.3, 0.5, 1.0, 5, 10 and 20 mL/h), two purging techniques (manually or infusion pump’s electronic bolus). Data were analyzed according to mean, standard deviation, Student’s t and ANOVA tests (p<0.05). Results: The start-up delay was greater in low rates regardless the priming technique. The electronic bolus increased the infusion pump A accuracy at 0.3mL/h (p=0.010), 0.5 mL/h (p=0.002) and 1.0mL/h (p=0.004). Pump’s accuracy in all studied rates and manual IV sets filling was similar. Conclusion: In low infusion rates the start-up delay was greater despite the infusion pump brand and electronic bolus improved pumps accuracy.

Publisher

FapUNIFESP (SciELO)

Subject

General Medicine

Reference15 articles.

1. Infusion Pumps,2018

2. Understanding infusion pumps;Mandel JE;Anesth Analg,2018

3. Hemolysis risk after packed red blood cells transfusion with infusion pumps;Wilson AMMM;Rev Lat-Am Enfermagem,2018

4. Intravenous smart pumps: usability issues, intravenous medication, administration error, and patient safety;Giuliano KK;Crit Care Nurs Clin North Am,2018

5. Use of technlogies in intravenous therapy: contributions to a safer practice;Moreira APA;Rev Bras Enferm,2017