Affiliation:
1. University of São Paulo
Abstract
The objective of the present study was to determine the reproducibility of the pixel values obtained with the Digora system (Soredex, Finland). Exposures were standardized, with variation in exposure and scanning time of two photostimulable phosphor plates containing a stepwedge image. The smallest variation in pixel values ranged from 50 to 75%, with the widest variations being observed in less dense steps. A significant difference in pixel values was observed in terms of X-ray exposure and scanning times and between the two plates themselves (ANOVA, p < 0.01). Using the present methodology, the reproducibility of pixel values was not satisfactory for the tested white photostimulable plates. This wide variation in digitalization might be influenced by the amount of X-rays that sensitized the plates. It may be important to establish the reproducibility of the pixel values in quantitative studies using digital image.
Subject
Electrical and Electronic Engineering,Atomic and Molecular Physics, and Optics
Cited by
5 articles.