Experimental assessment of three electrosurgical tissue-sealing devices in a porcine model

Author:

Dunay Miklós Pál1ORCID,Lipcsey Zsuzsanna1,Arany-Tóth Attila1,Németh Tibor1,Solymosi Norbert2,Venczel László3,Nagy Enikő3,Pap-Szekeres József3

Affiliation:

1. 1Department and Clinic of Surgery and Ophthalmology, University of Veterinary Medicine, István u. 2, H-1078, Budapest, Hungary

2. 2Centre of Bioinformatics, University of Veterinary Medicine, Budapest, Hungary

3. 3Department of General Surgery, County Teaching Hospital, Kecskemét, Hungary

Abstract

AbstractThree electrosurgical tissue-sealing devices (EnSeal ETSDRC-01, LigaSure LS1500 and Thunderbeat TB-0535PC) were compared regarding sealing time (ST), maximum working temperature (WTmax) and the total (MTZtotal) as well as the collateral microscopic thermal injury zone (MTZcollat) using laparoscopic handpieces 5 mm in diameter on four types of tissue (liver, mesentery, cross striated muscle and spleen) in an in vivo porcine model. LigaSure had the lowest mean ST in spleen, mesentery, muscle and liver, followed by Thunderbeat and EnSeal with significant differences between all types of tissues and devices. The significantly lowest mean WTmax was obtained for EnSeal in mesentery, muscle and liver. LigaSure and EnSeal operated at the lowest temperature in spleen without a significant difference between them. Thunderbeat produced significantly higher temperature peaks in all cases. The lowest mean MTZtotal was caused by LigaSure and EnSeal in spleen, mesentery and muscle without significant differences between them, followed by the significantly higher values of Thunderbeat. Nevertheless, Thunderbeat produced the significantly lowest mean MTZtotal in the liver. EnSeal produced the lowest mean MTZcollat in the liver, followed by LigaSure and Thunderbeat showing significant differences. EnSeal and LigaSure produced the lowest mean MTZcollat in the spleen, mesentery and muscle without significant differences between them, followed by the significantly higher values of Thunderbeat. Based on the results of this study, Thunderbeat seems to be more invasive to tissue integrity (even without the activation of the ultrasonic scissor function) than EnSeal or LigaSure, that operate at lower temperatures and were found to cause negligible collateral thermal damage.

Publisher

Akademiai Kiado Zrt.

Subject

General Veterinary

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3