Affiliation:
1. 1 Department of Cardiology, Zydus Hospital and Research Centre, Anand, India
Abstract
Failure to deflate the angioplasty balloon following its inflation during percutaneous coronary intervention (PCI) is an unusual complication. We report a case of an undeflatable stent balloon in a 70-year-old male, who underwent PCI for the right coronary artery after failed thrombolysis with streptokinase. After many failed attempts, the patient was cautiously managed by controlled puncture with a stiff guide wire. The procedure was successful without any hemodynamic compromise and ventricular tachycardia.