Author:
Filiaggi Mark J.,Coombs N. A.,Pilliar R.M.
Abstract
AbstractPlasma sprayed Hydroxyapatite (HA) coatings are applied to metal prostheses to allow for implant fixation through chemical bonding of the coating with surrounding bone tissue. Without a well-adhering coating, this fixation is threatened. Thus, a thorough characterization of the metal / ceramic interface is necessary. This study used a novel composite short bar interfacial fracture toughness technique with high resolution electron spectroscopic imaging to examine Ti-6AI-4V plasma spray coated with 100μm of HA. For this system, an interfacial fracture toughness value of 1.31 +/− 0.08 MPa·m1/2 was obtained, with a corresponding tensile adhesive bond strength of 6.7 +/− 1.5 MPa. High resolution ESI revealed distinct phosphorous segregation to the interface and diffusion into the underlying titanium. A 24-hour post-heat treatment at 960°C greatly increased the bond strength at this interface. Observations from ESI suggested that this effect may be due to enhanced diffusion of both phosphorous and calcium into the metal substrate.
Publisher
Springer Science and Business Media LLC
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献