Comparison of Transmittance and Reflectance Pulse Oximetry in Anesthetized Dogs

Author:

Nixdorff Jan,Zablotski Yury,Hartmann Katrin,Dörfelt Rene

Abstract

Objectives: The tongue is the standard site for placement of a pulse oximeter probe but is difficult to access during certain procedures such as dental and ophthalmic procedures and computerized tomography of the head. The aim of this study was to evaluate the performance of a new-generation reflectance pulse oximeter using the tail and tibia as sites for probe attachment.Materials and Methods: A total of 100 client-owned dogs that underwent anesthesia for various reasons were premedicated with butorphanol (n = 50; 0.2 mg/kg; group BUT) or butorphanol and dexmedetomidine (n = 50; 5 μg/kg; group DEX), administered intravenously. Anesthesia was induced with propofol and maintained with sevoflurane. A transmittance pulse oximeter probe was placed on the tongue and served as the reference standard. A reflectance probe was randomly placed on the tail base or the proximal tibia, and the position changed after testing. Signals from three consecutive measurements were obtained at each position. Failure was defined as “no signal,” “low signal,” or a pulse difference >10/min compared with the ECG heart rate. Data were analyzed using chi-square test, Wilcoxon matched-pair signed-rank test, and Bland-Altman analysis. P < 0.05 was considered significant.Results: In both groups (BUT and DEX), failure rate was higher when the tibia and tail were used as probe sites compared with the tongue. In both groups, the failure rate was higher for the tibia than for the tail. Dexmedetomidine-induced vasoconstriction increased failure rate at all probe positions.Clinical Significance: The tail base, but not the tibia, is an acceptable position for reflectance pulse oximeter probes in dogs. The tongue remains the probe site of choice, if accessible.

Publisher

Frontiers Media SA

Subject

General Veterinary

Reference24 articles.

1. The Australian incident monitoring study. Which monitor? An analysis of 2000 incident reports;Webb;Anaesth Intensive Care.,1993

2. Risk factors for anaesthetic-related death in cats: results from the confidential enquiry into perioperative small animal fatalities (CEPSAF);Brodbelt;Br J Anaesth.,2007

3. Pulse oximetry: principles and limitations;Sinex;Am J Emerg Med.,1999

4. Recent findings in the use of reflectance oximetry: a critical review;Keogh;Curr Opin Anaesthesiol.,2005

5. Design and evaluation of a new reflectance pulse oximeter sensor;Mendelson;Med Instrum.,1988

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Intranasal atomized dexmedetomidine alone or in combination with ketamine or midazolam to sedate healthy dogs;The Veterinary Journal;2024-10

2. The utility of the respiratory rate-oxygenation index as a predictor of treatment response in dogs receiving high-flow nasal cannula oxygen therapy;Frontiers in Veterinary Science;2024-05-07

3. Effects of Midazolam/Dexmedetomidine with Buprenorphine or Extended-release Buprenorphine Anesthesia in C57BL/6 Mice;Journal of the American Association for Laboratory Animal Science;2024-03-01

4. SPO2 Monitoring System;2023 International Conference on Computer Science and Emerging Technologies (CSET);2023-10-10

5. Nursing strategies for the mechanically ventilated patient;Frontiers in Veterinary Science;2023-07-28