Author:
Choi Dong Hyeok,Ahn So Hyun,Park Kwangwoo,Choi Sang Hyun,Kim Jin Sung
Abstract
PurposeThe purpose of this study is to propose a technique for delivering accurate doses in an image-guided system by developing an experimental setup optimized for total lymphoid irradiation (TLI) in rat lung transplantation.Materials and MethodsIn this study, a position-controlled shielding system was developed, and the dose was quantitatively evaluated using a 3D rat phantom and Gafchromic EBT3 film. In addition, we made our own image-guided system that allows the position of the rat and the shielding system to be confirmed during TLI.ResultsAs a result of using the position-controlled shielding system, it was found that the doses to the head and lungs were reduced by 93.1 and 87.4%, respectively, of the prescribed doses. In addition, it was shown that the position of the shielding system can be easily confirmed by using the image guidance system.ConclusionA shielding apparatus that can control dose delivery according to the size of the rat can optimize the dose for TLI in rat lung transplantation.