Li Chunqian,Zhao Shibin,Zhang Chenglun,Li Meng,Guo Jinjia,Dimova Natasha T.,Yang Tong,Liu Wen,Chen Guangquan,Yu Huaming,Xu Bochao
Abstract
Radon is an excellent natural tracer for studying various geophysical processes. In the past centuries, radon isotopes measurement approaches for marine research have been fully developed but still suffer limitations. Here we present the setup and validation of an improved continuous online measurement system (PIC-ORn) to measure dissolved radon in the surface ocean and other water bodies. We demonstrated that the PIC measurement efficiency is ~2 times higher than a RAD7 and is less affected by relative humidity and produces reliable results. Laboratorial measurements indicated that the new PIC-ORn system responded timely to the change of radon activities in water. The new system was successfully deployed during a cruise to the northwest Pacific Ocean in June 2021. Despite low radon-in-water activities, the results obtained by the new PIC-ORn system matched the traditional measurement systems within the estimated uncertainties. The PIC-ORn detector takes advantage of higher efficiency, lower cost, and power consumption, and is less affected by air moisture. The new system does not rely on drying units, further reducing on-site supervision, which would benefit the researches in submarine groundwater advection and diffusion and ocean-atmosphere gas exchange.
Funder
National Natural Science Foundation of China
Fundamental Research Funds for the Central Universities
Subject
Ocean Engineering,Water Science and Technology,Aquatic Science,Global and Planetary Change,Oceanography
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献