Characteristics analysis of compact cesium atomic clock with magnetic state selection
-
Published:2022-09-02
Issue:
Volume:10
Page:
-
ISSN:2296-424X
-
Container-title:Frontiers in Physics
-
language:
-
Short-container-title:Front. Phys.
Author:
Chen Jiang,Wang Ji,Guo Lei,Yang Jun,Ma Pei,Huang Liangyu,Liu Zhidong
Abstract
This paper introduces a compact magnetic state-selection cesium atomic clock called LIP Cs-3000 and analyses its characteristics. The test results reveal that the ranges of the line width, peak-to-valley ratio and signal-to-noise ratio of Ramsey pattern are 340–410 Hz, 2–13 and 2,000–8,000 respectively. The corresponding distributions of these parameters are derived by using statistical methods. According to these results, some suggestions are put forward for the further development of the clock. It is also pointed out that the most important factor affecting the accuracy of LIP Cs-3000 atomic clock is the uncertainty of cavity phase shift. In addition, the method to estimate the lifetime of a clock has been proposed. In the end, the paper gives some environmental adaptability designs of the clock, which enable the clock to be used in complex environments.
Funder
National Natural Science Foundation of China
Publisher
Frontiers Media SA
Subject
Physical and Theoretical Chemistry,General Physics and Astronomy,Mathematical Physics,Materials Science (miscellaneous),Biophysics
Reference23 articles.
1. The hydrogen maser and cesium clocks in time keeping at NTSC;Yuan,2009
2. Kalman filter characterization of cesium clocks and hydrogen masers;Breakiron,2002
3. GPS block IIF atomic frequency standard analysis;Vannicola,2010
4. GPS block IIF atomic frequency standard analysis;Vannicola,2013