Author:
Ma Fang-Zhou,Liu De-Feng,Yang An-Chao,Zhang Kai,Meng Fan-Gang,Zhang Jian-Guo,Liu Huan-Guang
Abstract
IntroductionThis work aims to assess the accuracy of robotic assistance guided by a videometric tracker in deep brain stimulation (DBS).MethodsWe retrospectively reviewed a total of 30 DBS electrode implantations, assisted by the Remebot robotic system, with a novel frameless videometric registration workflow. Then we selected 30 PD patients who used stereotactic frame surgery to implant electrodes during the same period. For each electrode, accuracy was assessed using radial and axial error.ResultsThe average radial error of the robot-assisted electrode implantation was 1.28 ± 0.36 mm, and the average axial error was 1.20 ± 0.40 mm. No deaths or associated hemorrhages, infections or poor incision healing occurred.ConclusionRobot-assisted implantation guided by a videometric tracker is accurate and safe.
Funder
Beijing Municipal Science and Technology Commission
Subject
Artificial Intelligence,Biomedical Engineering
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献