Performance Characterization of Dosimeters Based on Radiation-Hard Silicon Diodes in Gamma Radiation Processing

Author:

Bueno C. C.,Camargo F.,Gonçalves J. A. C.,Pascoalino K.,Mangiarotti A.,Tuominen E.,Härkönen J.

Abstract

The dosimetric response of silicon diodes produced with distinct engineering technologies, Magnetic Czochralski (MCz), and standard Float Zone (Fz), has been investigated, aiming at their use for online dosimeters in gamma radiation processing applications. The p+-n-n+ junction diodes, 300 µm thick with an active area of 25 mm2, are operated as online radiation dosimeters in the short-circuit current mode. In this case, the key dosimetric quantity is the dose rate, which is correlated with the output current from the diode subjected to radiation. Thus, the dose is obtained offline by the integration of the corresponding current signal. The irradiations are performed with an industrial Gammacell 60Co facility at 2.3–2.44 kGy/h covering doses up to 275 kGy. Under continuous irradiation, both diodes delivered current signals whose intensities decreased with accumulated doses. Mitigation of this decay has been accomplished by pre-irradiating the devices to 700 kGy. Polynomial functions best represent the dose responses for either pristine or preirradiated diodes. The relevant dosimetric parameters as response stability, charge sensitivity, and repeatability of current signals (<5%) reveal the better performance of the MCz diode. It is important to note that the whole dataset fully complies with the international standard protocols for routine dosimeters in radiation processing dosimetry. Regarding radiation damage, which in unbiased diodes manifests primarily in the decay of current sensitivity, the results also showed greater tolerance of the MCz diode. Based on these studies, large availability, and better cost-effectiveness, it is possible to endorse the potential use of MCz devices as online routine dosimeters in radiation processing applications. However, the data reproducibility with the accumulated dose, the dose lifespan, and the effect of the irradiation conditions (e.g., temperature, relative humidity, and dose fractionating) remain to be investigated. Works in this direction are currently in progress.

Funder

Conselho Nacional de Desenvolvimento Científico e Tecnológico

Publisher

Frontiers Media SA

Reference37 articles.

1. Feasibility Study of a Current Mode Gamma Radiation Dosimeter Based on a Commercial PIN Photodiode and a Custom Made Auto-Ranging Electrometer;Andjelkovic;Nucl. Tech. Rad Prot.,2013

2. Measuring and Interpreting the Lifetime of Silicon Wafers;Cuevas;Solar Energy,2004

3. Silicon Diode Dosimetry;Dixon;Int. J. Appl. Radiat. Isotopes,1982

4. Radiation Damage Studies on MCz and Standard and Oxygen Enriched Epitaxial Silicon Devices;Fretwurst;Nucl. Instr. Methods Phys. Res. Section A: Acc. Spectrometers, Detectors Associated Equipment,2007