Author:
Luo Li,Qin Zuoan,Shi Ruizheng,Ge Liangqing
Abstract
Atrial tachycardia originating from the right atrial appendage has a higher probability of failure of catheter ablation. Here we report a case of a 13-year-old boy with incessant tachycardia, complicated by heart enlargement, and heart failure. Electrophysiological examination showed that atrial tachycardia (AT) originated from the apex of the right atrial appendage, and endocardial catheter ablation was ineffective. After thoracoscopic approach, the right atrial appendage was successfully ablated with bipolar radiofrequency ablation forceps, atrial tachycardia was terminated and sinus rhythm was restored. Within 3 months since the patient was discharged from the hospital, no arrhythmia occurred and the heart structure returned to normal. Thus, thoracoscopic clamp radiofrequency ablation may be a reasonable choice for young patients with atrial tachycardia originated from the right atrial appendage when transendocardial ablation is not effective.
Subject
Cardiology and Cardiovascular Medicine
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献