Positron emission tomography PET/CT harmonisation study of different clinical PET/CT scanners using commercially available software

Author:

Lowe Gerry1,Spottiswoode Bruce2,Declerck Jerome2,Sullivan Keith3,Sharif Mhd Saeed4,Wong Wai-Lup5,Sanghera Bal5

Affiliation:

1. Cancer Centre, Mount Vernon Hospital, London, UK

2. Siemens Medical Solutions USA, Inc., Knoxville, TN, USA

3. Health Research Methods Unit, University of Hertfordshire, Hertfordshire, UK

4. School of Arch., Comp. and Eng., University of East London, London, UK

5. Paul Strickland Scanner Centre, Mount Vernon Hospital, London, UK

Abstract

Objectives: Harmonisation is the process whereby standardised uptake values from different scanners can be made comparable. This PET/CT pilot study aimed to evaluate the effectiveness of harmonisation of a modern scanner with image reconstruction incorporating resolution recovery (RR) with another vendor older scanner operated in two-dimensional (2D) mode, and for both against a European standard (EARL). The vendor-proprietary software EQ•PET was used, which achieves harmonisation with a Gaussian smoothing. A substudy investigated effect of RR on harmonisation. Methods: Phantom studies on each scanner were performed to optimise the smoothing parameters required to achieve successful harmonisation. 80 patients were retrospectively selected; half were imaged on each scanner. As proof of principle, a cohort of 10 patients was selected from the modern scanner subjects to study the effects of RR on harmonisation. Results: Before harmonisation, the modern scanner without RR adhered to EARL specification. Using the phantom data, filters were derived for optimal harmonisation between scanners and with and without RR as applicable, to the EARL standard. The 80-patient cohort did not reveal any statistically significant differences. In the 10-patient cohort SUVmax for RR > no RR irrespective of harmonisation but differences lacked statistical significance (one-way ANOVA F(3.36) = 0.37, p = 0.78). Bland-Altman analysis showed that harmonisation reduced the SUVmax ratio between RR and no RR to 1.07 (95% CI 0.96–1.18) with no outliers. Conclusions: EQ•PET successfully enabled harmonisation between modern and older scanners and against the EARL standard. Harmonisation reduces SUVmax and dependence on the use of RR in the modern scanner. Advances in knowledge: EQ•PET is feasible to harmonise different PET/CT scanners and reduces the effect of RR on SUVmax.

Publisher

British Institute of Radiology

Subject

Materials Chemistry,Economics and Econometrics,Media Technology,Forestry

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3