Occupational and patient doses for interventional radiology integrated into a dose management system

Author:

Vano Eliseo12ORCID,Fernandez-Soto José M1,Ten José I1,Sanchez Casanueva Roberto M12

Affiliation:

1. Instituto de Investigación Sanitaria del Hospital Clínico San Carlos and Medical Physics Service, Madrid, Spain

2. Department of Radiology, Medical School. Complutense University, Madrid, Spain

Abstract

Objectives: The International Commission on Radiological Protection recommends managing patient and occupational doses as an integrated approach, for the optimisation of interventional procedures. The conventional passive personal dosimeters only allow one to know the accumulated occupational doses during a certain period of time. This information is not enough to identify if there is a lack of occupational radiation protection during some procedures. This paper describes the use of a dose management system (DMS) allowing patient and occupational doses for individual procedures to be audited. Methods: The DMS manages patient and occupational doses measured by electronic personal dosimeters. One dosemeter located at the C-arm is used as a reference for scatter radiation. Data have been collected from five interventional rooms. Dosimetry data can be managed for the whole procedure and the different radiation events. Optimisation is done through auditing different sets of parameters for individual procedures: patient dose indicators, occupational dose values, the ratio between occupational doses, and the doses measured by the reference dosemeter at the C-arm, and the ratio between occupational and patient dose values. Results: The managed data correspond to the year 2021, with around 4500 procedures, and 8000 records on occupational exposures. Patient and staff dose data (for 11 cardiologists, 7 radiologists and 8 nurses) were available for 3043 procedures. The DMS allows alerts for patient dose indicators and occupational exposures to be set. Conclusions: The main advantage of this integrated approach is the capacity to improve radiation safety for patients and workers together, auditing alerts for individual procedures. Advances in knowledge: The management of patient and occupational doses together (measured with electronic personal dosimeters) for individual interventional procedures, using dose management systems, allows alerting optimisation on high-dose values for patients and staff.

Publisher

British Institute of Radiology

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3