Image-mode performance characterisation of a positron emission tomography subsystem designed for Biology-guided radiotherapy (BgRT)

Author:

Hu Zhiqiang1,Bieniosek Matthew1,Ferri Valentina2,Iagaru Andrei2,Kovalchuk Nataliya3,Han Bin3,Xing Lei3,Vitzthum Lucas3,Olcott Peter1,Narayanan Manoj1,Laurence Thomas1,Ren Yulan3,Oderinde Oluwaseyi M1,Shirvani Shervin M1,Chang Daniel3,Surucu Murat3ORCID

Affiliation:

1. RefleXion Medical, Inc, Hayward, CA

2. Department of Radiology, Stanford University, Stanford, CA

3. Department of Radiation Oncology, Stanford University, Stanford, CA

Abstract

Objectives: In this study, we characterise the imaging-mode performance of the positron emission tomography (PET) subsystem of the RefleXion X1 machine using the NEMA NU-2 2018 standard. Methods: The X1 machine consists of two symmetrically opposing 900 arcs of PET detectors incorporated into the architecture of a ring-gantry linear accelerator rotating up to 60 RPM. PET emissions from a tumour are detected by the PET detectors and used to guide the delivery of radiation beam. Imaging performance of the PET subsystem on X1 machine was evaluated based on sensitivity of the PET detectors, spatial resolution, count-loss performance, image quality, and daily system performance check. Results: PET subsystem sensitivity was measured as 0.183 and 0.161 cps/kBq at the center and off-center positions, respectively. Spatial resolution: average FWHM values of 4.3, 5.1, and 6.7 mm for the point sources at 1, 10, and 20 cm off center, respectively were recorded. For count loss, max NECR: 2.63 kcps, max true coincidence rate: 5.56 kcps, and scatter fraction: 39.8%. The 10 mm sphere was not visible. Image-quality contrast values were: 29.6%, 64.9%, 66.5%, 81.8%, 81.2%, and background variability: 14.8%, 12.4%, 10.3%, 8.8%, 8.3%, for the 13, 17, 22, 28, 37 mm sphere sizes, respectively. Conclusions: When operating in an imaging mode, the spatial resolution and image contrast of the X1 PET subsystem were comparable to those of typical diagnostic imaging systems for large spheres, while the sensitivity and count rate were lower due to the significantly smaller PET detector area in the X1 system. Clinical efficacy when used in BgRT remains to be validated. Advances in knowledge: This is the first performance evaluation of the PET subsystem on the novel BgRT machine. The dual arcs rotating PET subsystem on RefleXion X1 machine performance is comparable to those of the typical diagnostic PET system based on the spatial resolution and image contrast for larger spheres.

Publisher

British Institute of Radiology

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3