Does the shape of the field-of-view influence the magnitude of artefacts from high-density materials in cone-beam computed tomography?

Author:

Cascante-Sequeira Deivi1,Fontenele Rocharles Cavalcante12,Martins Luciano Augusto Cano1,Brasil Danieli Moura3,Oliveira Matheus L1,Freitas Deborah Queiroz1,Haiter-Neto Francisco1

Affiliation:

1. Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas , Piracicaba, Brazil

2. Department of Imaging and Pathology, OMFS IMPATH Research Group, Faculty of Medicine, KU Leuven , Leuven, Belgium

3. Department of Oral Diagnosis & Oral Health, University of Louisville School of Dentistry , Louisville, United States of America

Abstract

Objective: To compare cylindrical and convex triangular field-of-views (FOVs) concerning the magnitude of artefacts from high-density materials in cone-beam CT (CBCT). Methods and materials: Cylinders of amalgam, chromium-cobalt, titanium, and zirconia were individually placed in the anterior and posterior regions of a polymethylmethacrylate phantom and scanned using cylindrical and convex triangular FOVs of the Veraview X800 CBCT device. Using the Image J software, 15 square regions of interest (ROIs) were placed in the axial reconstruction around the middle level of the cylinder and at distances of 0.5, 1.0, and 1.5 cm from the centre of the cylinder. Mean grey value and standard deviation of each ROI were averaged for each distance and subtracted from the values of a control ROI to calculate the magnitude of the artefacts by the grey value mean difference (GVMD) and grey value standard deviation (GVSD). Multiway analysis of variance with Tukey post-hoc test with a significance level of 5% evaluated the effect of the shape of the FOV, position inside the FOV, high-density material, and the distance of the artefact from the material. Results: The convex triangular FOV increased the GVSD for all materials in the anterior and posterior regions at 0.5 cm compared to the cylindrical FOV (p < 0.0001). The convex triangular FOV showed greater GVMD for chromium-cobalt and zirconium in the anterior region and all materials in the posterior region at all distances (p < 0.0001). Conclusion: The FOV shape influences the magnitude of artefacts from high-density materials. The convex triangular FOV showed greater artefact magnitude with variability among the high-density materials, region in the FOV, and distance from the material.

Publisher

Oxford University Press (OUP)

Subject

General Dentistry,Radiology, Nuclear Medicine and imaging,General Medicine,Otorhinolaryngology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3