Monitoring cone-beam CT radiation dose levels in a University Hospital

Author:

Brasil Danieli Moura12ORCID,Merken Karen3,Binst Joke4,Bosmans Hilde34,Haiter-Neto Francisco1,Jacobs Reinhilde25ORCID

Affiliation:

1. Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, SP, Brazil

2. OMFS-IMPATH Research Group, Department of Imaging & Pathology, Faculty of Medicine, KU Leuven and Department of Oral and Maxillofacial Surgery, Campus Sint Rafaël, University Hospitals Leuven, Leuven, Belgium

3. Department of Imaging and Pathology, KU Leuven, Division of Medical Physics & Quality Assessment, Leuven, Belgium

4. Department of Radiology, UZ Leuven, Leuven, Belgium

5. Department of Dental Medicine, Karolinska Institutet, Stockholm, Sweden

Abstract

Objective: To present patient dose levels for different CBCT scanners, acquired by a dose monitoring tool in a University Hospital, as a function of field of view (FOV), operation mode, and patient age. Methods: An integrated dose monitoring tool was used to collect radiation exposure data [type of CBCT unit, dose-area product (DAP), FOV size, and operation mode] and patient demographic information (age, referral department) performed on a 3D Accuitomo 170 and a Newtom VGI EVO unit. Effective dose conversion factors were calculated and implemented into the dose monitoring system. For each CBCT unit, the frequency of examinations, clinical indications, and effective dose levels were obtained for different age and FOV groups, and operation modes. Results: A total of 5163 CBCT examinations were analyzed. Surgical planning and follow-up were the most frequent clinical indications. For the standard operation mode, effective doses ranged from 35.1 to 300 µSv and 9.26-117 µSv using 3D Accuitomo 170 and Newtom VGI EVO, respectively. In general, effective doses decreased with increasing age and FOV size reduction. Conclusions: Effective dose levels varied notably between systems and operation modes.Operation mode selection and FOV size were indication-oriented, with larger FOV sizes election serving surgical planning and follow-up. Seeing the influence of FOV size on effective dose levels, manufacturers could be advised to move toward patient-specific collimation and dynamic FOV selection. Systematically monitoring patient doses could be recommended for steering future CBCT optimization.

Publisher

Oxford University Press (OUP)

Subject

General Dentistry,Radiology, Nuclear Medicine and imaging,General Medicine,Otorhinolaryngology