Comparison of subjective image analysis and effective dose between low-dose cone-beam computed tomography machines

Author:

Charuakkra Arnon1,Mahasantipiya Phattaranant1,Lehtinen Antti2,Koivisto Juha3,Järnstedt Jorma2

Affiliation:

1. Division of Oral and Maxillofacial Radiology, Department of Oral Biology and Diagnostic Sciences, Faculty of Dentistry, Chiang Mai University, Chiang Mai, Thailand

2. Medical Imaging Center, Department of Radiology, Tampere University Hospital, Tampere, Finland

3. Department of Physics, University of Helsinki, Helsinki, Finland

Abstract

Objectives: To compare the cone-beam computed tomography (CBCT) image quality and effective dose between low-dose scanning and standard manufacturer-recommended protocols among different CBCT units. Methods: Three human-equivalent phantoms were scanned using the ultra-low-dose (ULD), low dose (LD), and standard dose (STD) modes of ProMax 3D Mid (Planmeca Oy, Helsinki, Finland) and Orthophos SL (Sirona, Bensheim, German) for the CBCT images. The quality of the dental anatomical images was assessed by four experienced oral and maxillofacial radiologists using a 5-point Likert scale. OnDemand3D (Cybermed Co., Seoul, Korea) was used as the third-party software for viewing. The percentage of absolute agreement was calculated to determine intra- and interrater agreements among the observers. The effective doses for all CBCT scanning protocols were also calculated. Results: The STD protocol yielded a higher image quality than did the ULD and LD protocols in both ProMax 3D Mid and Orthophos SL. The ULD and LD protocols demonstrated an “acceptable-to-good” sense of visual perception of the CBCT images. The visibility scores significantly differed between the ULD and LD and the STD protocols in ProMax 3D Mid and Orthophos SL, except for the 120-kVp setting in ProMax 3D Mid. The average intra- and interrater agreement scores ranged from 0.63 to 0.89 and from 0.44 to 0.76, respectively. The ULD and LD protocols reduced the radiation dose sixfold compared with the STD protocol. Conclusions: High-tube-voltage protocols could remarkably reduce the imaging dose without degrading the image quality. Specifically, ULD and LD CBCT protocols may be adopted as routine practice for diagnosis and treatment planning.

Publisher

British Institute of Radiology

Subject

General Dentistry,Radiology, Nuclear Medicine and imaging,General Medicine,Otorhinolaryngology

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3