Interobserver reproducibility of contrast-enhanced ultrasound in diabetic nephropathy

Author:

Lin Lin1,Wang Yiru1,Yan Lin1,Li Nan1,Tian Xiaoqi1,Li Qiuyang1,Tang Jie1,Luo Yukun1

Affiliation:

1. Department of Ultrasound, The First Medical Center, Chinese PLA General Hospital, Beijing, China

Abstract

Objective: To investigate the interobserver reproducibility of contrast-enhanced ultrasound (CEUS) in patients with diabetic nephropathy. Methods: A total of 40 patients with diabetic nephropathy were enrolled in this retrospective study. We measured peak intensity (PEAK), time-to-peak (TP), area under the curve (AUC) and mean transit time (MTT) of renal CEUS in each patient. Each parameter was performed by two independent observers. The interobserver reproducibility was assessed using intraclass correlation coefficients (ICCs) with 95% confidence intervals (CIs) and Bland–Altman plots by mean difference with 95% limits of agreement (LOAs). Results: The parameters of the left and right kidneys showed moderate or good reliability. The best was the left kidney AUC parameters (ICC,0.945),with a 95% CI of 0.896–0.971.The Bland–Altman plots showed that the mean differences between the right renal parameters (PEAK, TP, AUC and MTT) obtained between the observers 1 and 2 were −6.63%,2.54%,−11.30%,−2.22%, and the 95% LOAs are −30.78 to 17.52%, −30.15 to 35.23%, −43.95 to 21.35%, −44.50 to 40.06%. While the left ones were −2.89%, 0.32%, −8.26%, 1.25% , and the 95% LOAs were −38.50 to 32.72%, −31.98 to 32.62%, −57.89 to 41.37%, −37.21 to 39.71%, respectively. Conclusion: Quantitative CEUS can show good interobserver reproducibility, which is better for the right kidney. It seems necessary to establish standardized techniques for obtaining contrast-enhanced quantitative analysis of renal blood perfusion. Advances in knowledge: Quantitative assessment of renal perfusion by CEUS in patients with diabetic nephropathy provides diagnostic information. Furthermore, renal perfusion assessment in patients with diabetic nephropathy using CEUS has good reproducibility.

Publisher

British Institute of Radiology

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3