Retrospective analysis of reduced energy switching and room switching times on throughput efficiency of a multi-room proton therapy center

Author:

Mah Dennis1,Chen Chin Cheng1,Nawaz A Omer1,Galbreath Greg2,Shmulenson Reuven1,Lee Nancy3,Chon Brian1

Affiliation:

1. ProCure Proton Therapy Center, 103 Cedar Grove Lane, Somerset NJ 08540, USA

2. Transeo Radiothearpy Solutions, 1 Ferry Building #255, San Francisco CA 94111, USA

3. Memorial Sloan Kettering Cancer Center, 1275 York Avenue, New York, NY 10065, USA

Abstract

Objective: To quantify how a control software upgrade changed beam delivery times and impacted efficiency and capacity of a multiroom proton therapy center. Methods: A four-room center treating approximately 90 patients/day, treating for approximately 7 years with optimized operations, underwent a software upgrade which reduced room and energy switching times from approximately 30 to 20 s and approximately 4 s to ~0.5 s, respectively. The center uses radio-frequency identification data to track patient treatments and has software which links this to beam delivery data extracted from the treatment log server. Two 4-month periods, with comparable patient volume, representing periods before and after the software change, were retrospectively analyzed. Results: A total of 16,168 and 17,102 fields were analyzed. For bilateral head and neck and prostate patients, the beam waiting time was reduced by nearly a factor of 3 and the beam delivery times were reduced by nearly a factor of 2.5. Room switching times were reduced more modestly. Gantry capacity has increased from approximately 30 patients to 40–45 patients in a 16-h daily operation. Conclusions: Many proton centers are striving for increased efficiencies. We demonstrated that reductions in energy and room switching time can significantly increase center capacity. Greater potential for further gains would come from improvements in setup and imaging efficiency. Advances in knowledge: This paper provides detailed measured data on the effect on treatment times resulting from reducing energy and room switching times under controlled conditions. It helps validate the models of previous investigations to establish treatment capacity of a proton therapy center.

Publisher

British Institute of Radiology

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3