Establishing how patient size and degree of miscentring affect CTDIvol, using patient data from a dose tracking system

Author:

Cournane Sean1,Brunell Eimear2,Rowan Michael3

Affiliation:

1. Department of Medical Physics and Bioengineering, St James’s hospital, Dublin, Ireland

2. Department Medical Physics and Clinical engineering, Tallaght University Hospital, Dublin, Ireland

3. Department of Radiology, Mater Misericordiae Hospital, Dublin, Ireland

Abstract

Objectives: The work aimed to investigate and model the relationship between CT dose index (CTDIvol), patient size and miscentring metrics using data extracted from a dose tracking software tool. Further, using the established model we aimed to extract and estimate further AEC metrics and, finally, demonstrate how the CTDIvol may vary across a range of imaging exams and with variations in patient size, positional metrics and with scan parameters. Methods: CT scan parameters, patient dose metrics, patient positioning information and patient and water equivalent diameter information was extracted from abdomen, thorax–abdomen–pelvis and thorax imaging exams acquired on a 128-slice Siemens Somatom Definition AS + CT system using dose tracking software over a 1 year period. A miscentring factor, accounting for the magnification of the patient due to miscentring was derived. Significant predictors ( p-value < 0.001) of the CTDIvol were considered as inputs into the model following regression analysis. Results: The model was capable of describing the CTDIvol and its variation with patient miscentring and patient size. AEC parameters, such as the reference CTDIvol, reference diameter and the AEC strength were estimated. Further, the model allowed for comparisons of how changes in scan settings, such as kVp, pitch and slice thickness affected CTDIvol. Conclusions: We demonstrate the use of clinical data, extracted from dose tracking software, to investigate and monitor AEC behaviour and CT output. Advances in knowledge: The presented model allows for the remote investigation of AEC behaviour using patient data.

Publisher

British Institute of Radiology

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3