Evaluation of stray radiation to the operator for five hand-held dental X-ray devices

Author:

Smith Richard1,Tremblay Richard1,Wardlaw Graeme M1

Affiliation:

1. Consumer and Clinical Radiation Protection Bureau, Health Canada, Canada

Abstract

Objectives: Evaluate stray radiation to the operator, as represented by a plane within the significant zone of occupancy (SZO), produced by five models of hand-held intraoral dental X-ray devices (HIDXDs). Methods: The stray radiation for five models of HIDXDs was measured, using an anthropomorphic tissue-equivalent head phantom as a scattering object. An ionization chamber was used to measure the air kerma (μGy) at 63 positions in a 160 cm high by 60 cm wide plane that was 10 cm behind the X-ray device, identified as being within the SZO. Results: Based on the measured air kerma from stray radiation of five different HIDXDs, the estimated annual air kerma at all measured spatial positions was calculated. When calculated using a median air kerma of 0.8 mGy at the distal end of the cone, as typically required for digital image receptors, 1 the ranges for estimated annual air kerma in the SZO across the devices were 0.14–0.77 mGy for the median, 0.41–1.01 mGy for the mean, and 1.32–2.55 mGy for the maximum. Similarly, when calculated using a median air kerma of 1.6 mGy as typically required for D-speed film, 2 the ranges for estimated annual air kerma across the devices were 0.28–1.54 mGy for the median, 0.83–2.03 mGy for the mean, and 2.64–5.10 mGy for the maximum. Conclusions: From measured air kerma values of stray radiation in the SZO, estimated annual exposures to the operator for HIDXDs are expected to be greater than from conventional wall-mounted or portable devices activated from a protected area (at a distance or behind shielding). HIDXDs should therefore only be used when patient accessibility makes their use necessary and the use of a portable device on a stand or a wall-mounted device is not reasonably feasible. This approach would keep occupational radiation exposures of dental workers as low as reasonably achievable.

Publisher

British Institute of Radiology

Subject

General Dentistry,Radiology Nuclear Medicine and imaging,General Medicine,Otorhinolaryngology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3