Effect of micro-computed tomography reconstruction protocols on bone fractal dimension analysis

Author:

Gaêta-Araujo Hugo1,Oliveira-Santos Nicolly1,Brasil Danieli Moura1,do Nascimento Eduarda Helena Leandro1,Madlum Daniela Verardi1,Haiter-Neto Francisco1,Oliveira-Santos Christiano de2

Affiliation:

1. Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas (UNICAMP), Av. Limeira, Piracicaba, São Paulo, Brazil

2. Department of Stomatology, Public Oral Health, Forensic Dentistry, Division of Oral Radiology, School of Dentistry of Ribeirao Preto, University of Sao Paulo, Ribeirão Preto, São Paulo, Brazil

Abstract

Objectives: To evaluate the influence of the level of three micro-CT reconstruction tools: beam-hardening correction (BHC), smoothing filter (SF), and ring artefact correction (RAC) on the fractal dimension (FD) analysis of trabecular bone. Methods: Five Wistar rats’ maxillae were individually scanned in a SkyScan 1174 micro-CT device, under the following settings: 50 kV, 800 µA, 10.2 µm voxel size, 0.5 mm Al filter, rotation step 0.5°, two frames average, 180° rotation and scan time of 35 min. The raw images were reconstructed under the standard protocol (SP) recommended by the manufacturer, a protocol without any artefact correction tools (P0) and 35 additional protocols with different combinations of SF, RAC and BHC levels. The same volume of interest was established in all reconstructions for each maxilla and the FD was calculated using the Kolmogorov (box counting) method. One-way ANOVA with Dunnet’s post-hoc test was used to compare the FD of each reconstruction protocol (P0–P35) with the SP (α = 5%). Multiple linear regression verified the dependency of reconstruction tools in FD. Results: Overall, FD values are not dependent on RAC (p = 0.965), but increased significantly when the level of BHC and SF increased (p < 0.001). FD values from protocols with BHC at 45% combined with SF of 2, and BHC at 30% combined with SF of 4 or 6 had no statistical difference compared to SP. Conclusions: BHC and SF tools affect the FD values of micro-CT images of the trabecular bone. Therefore, these reconstruction parameters should be standardized when the FD is analyzed.

Publisher

British Institute of Radiology

Subject

General Dentistry,Radiology, Nuclear Medicine and imaging,General Medicine,Otorhinolaryngology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3