Relation of Crown Failure Load to Flexural Strength for Three Contemporary Dental Polymers

Author:

Alghazzawi Tariq F.12

Affiliation:

1. Department of Substitutive Dental Sciences, Taibah University, Madinah 42353, Saudi Arabia

2. Department of Mechanical and Materials Engineering, The University of Alabama at Birmingham, Birmingham, AL 35294, USA

Abstract

Polymeric materials show great promise for use in a variety of dental applications. Manufacturers generally provide flexural strength information based on standardized (ISO and ASTM) specimen dimensions and loading conditions. It is not clear, however, if flexural strength data are predictive of the clinical performance of dental crowns. The objectives of this study were, therefore, to determine whether flexural strengths, as measured via three-point bending (3PB), would be predictive of failure loads assessed via crunch-the-crown (CTC) tests. Three brands of polymers (Trilor, Juvora, and Pekkton) were fabricated into rectangular bars and fully contoured crowns (10 specimens of each polymer brand, 30 specimens of each shape). Differential scanning calorimetry (DSC), Fourier transform infrared spectroscopy (FTIR), and burn off tests were used to characterize/confirm the materials. Bars were tested blindly in 3PB to determine flexural strength, and crowns were CTC-tested to determine failure load after luting to resin abutments. The statistical significance of the test results was evaluated via one-way ANOVA (α = 0.05) and Pearson’s correlation coefficient, while regression analysis was used to test for a correlation between 3PB and CTC results. The fracture mechanisms and failure surface characteristics were characterized using scanning electron microscopy (SEM). There were significant differences (p < 0.05) in the mean crown failure loads (Trilor (7033 N) > Juvora (5217 N) > Pekkton (3023 N)) and mean flexural strengths of the bars (Trilor (468 MPa) > Juvora (197 MPa) = Pekkton (192 MPa)). The mode of crown fracture was different between the materials and included deformation (Juvora), ductile-to-brittle fracture (Pekkton), and a combination of cracks and deformation (Trilor). Flexural strengths did not correlate with the corresponding crown failure loads for any of the materials tested. These results suggest that dental practitioners should not rely on the flexural strengths reported from three-point bending tests, as advertised by the manufacturer, to predict the performance of polymeric crowns.

Funder

Council for International Exchange of Scholars

Publisher

MDPI AG

Subject

Polymers and Plastics,General Chemistry

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3