Abstract
Older adults walk with a shorter stride length, reduced hip range of motion (ROM) and higher cadence. These are signs of reductions in walking ability. This study investigated whether using a wireless smart insole system that monitored and provided biofeedback to encourage an extension of swing time could increase stride length and hip flexion, while reducing the cadence. Seven older adults were tested in this study, with and without the biofeedback device, in an outdoor environment. Gait analysis was performed by using GaitRite system and Xsens MVN. Repeated measures analysis demonstrated that with biofeedback, the swing time increased by 6.45%, stride length by 4.52% and hip flexion by 14.73%, with statistical significance. It also decreased the cadence significantly by 5.5%. This study has demonstrated that this smart insole system modified positively the studied gait parameters in older adults and has the potential to improve their walking ability.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献