Abstract
In this study, we developed a range of motion sensing system (ROMSS) to simulate the function of the elbow joint, with errors less than 0.76 degrees and 0.87 degrees in static and dynamic verification by the swinging and angle recognition modules, respectively. In the simulation process, the ɣ correlation coefficient of the Pearson difference between the ROMSS and the universal goniometer was 0.90, the standard deviations of the general goniometer measurements were between ±2 degrees and ±2.6 degrees, and the standard deviations between the ROMSS measurements were between ±0.5 degrees and ±1.6 degrees. With the ROMSS, a cloud database was also established; the data measured by the sensor could be uploaded to the cloud database in real-time to provide timely patient information for healthcare professionals. We also developed a mobile app for smartphones to enable patients and healthcare providers to easily trace the data in real-time. Historical data sets with joint activity angles could be retrieved to observe the progress or effectiveness of disease recovery so the quality of care could be properly assessed and maintained.
Funder
Chang Gung Memorial Hospital
National Taipei University
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献