Abstract
Herein, we present reliable, robust, stable, and cost-effective solid-contact ion-selective electrodes (ISEs) for perchlorate determination. Single-walled carbon nanotubes (SWCNTs) were used as solid-contact material and indium (III) 5, 10, 15, 20-(tetraphenyl) porphyrin chloride (InIII-porph) as an ion carrier. The sensor exhibited an improved sensitivity towards ClO4− ions with anionic slope of −56.0 ± 1.1 (R2 = 0.9998) mV/decade over a linear range 1.07 × 10−6 – 1.0 × 10−2 M and detection limit of 1.8 × 10−7 M. The short-term potential stability and the double-layer capacitance were measured by chronopotentiometric and electrochemical impedance spectroscopy (EIS) measurements, respectively. The sensor is used for ClO4− determination in fireworks and propellant powders. The results fairly agree with data obtained by ion chromatography.
Funder
Deanship of Scientific Research, King Saud University
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
16 articles.
订阅此论文施引文献