Abstract
Procalcitonin (PCT) is a known protein biomarker clinically used for the early stages of sepsis diagnosis and therapy guidance. For its reliable determination, sandwich format magnetic bead-based immunoassays with two different electrochemical detection approaches are described: (i) disposable screen-printed carbon electrodes (SPE-C, on-drop detection); (ii) electro-kinetically driven microfluidic chips with integrated Au electrodes (EMC-Au, on-chip detection). Both approaches exhibited enough sensitivity (limit of detection (LOD) of 0.1 and 0.04 ng mL−1 for SPE-C and EMC-Au, respectively; cutoff 0.5 ng mL−1), an adequate working range for the clinically relevant concentrations (0.5–1000 and 0.1–20 ng mL−1 for SPE-C and EMC-Au, respectively), and good precision (RSD < 9%), using low sample volumes (25 µL) with total assay times less than 20 min. The suitability of both approaches was successfully demonstrated by the analysis of human serum and plasma samples, for which good recoveries were obtained (89–120%). Furthermore, the EMC-Au approach enabled the easy automation of the process, constituting a reliable alternative diagnostic tool for on-site/bed-site clinical analysis.
Subject
Clinical Biochemistry,General Medicine
Cited by
31 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献