Abstract
Moses technology was born with the aim of controlling the Moses effect present in every single Ho:YAG laser lithotripsy. The capacity to divide the energy pulse into two sub-pulses gained popularity due to the fact that most of the energy would be delivered in the second pulse. However, is this pulse modulation technique really better for endocorporeal laser lithoripsy? A review of the literature was performed and all relevant clinical trials of Moses 1.0 and 2.0, as well as the lab studies of Moses 2.0 carried out up to June 2022 were selected. The search came back with 11 clinical experiences (10 full-text clinical trials and one peer-reviewed abstract) with Moses 1.0 and Moses 2.0, and three laboratory studies (peer-reviewed abstracts) with Moses 2.0 only. The clinical experiences confirmed that the MT (1.0) has a shorter lasing time but lower laser efficacy, because it consumes more J/mm3 when compared with the LP Ho:YAG laser (35 W). This gain in lasing time did not provide enough savings for the medical center. Additionally, in most comparative studies of MT (1.0) vs. the regular mode of the HP Ho:YAG laser, the MT did not have a significant different lasing time, operative time or stone-free rate. Clinical trials with Moses 2.0 are lacking. From what has been published until now, the use of higher frequencies (up to 120 Hz) consumes more total energy and J/mm3 than Moses 1.0 for similar stone-free rates. Given the current evidence that we have, there are no high-quality studies that support the use of HP Ho:YAG lasers with MT over other lasers, such as LP Ho:YAG lasers or TFL lasers.
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献