Affiliation:
1. School of Electrical Engineering and Computer Science, Gwangju Institute of Science and Technology, 123 Cheomdangwagi-ro, Buk-gu, Gwangju 61005, Republic of Korea
Abstract
A method to enhance laser ultrasound (LUS) image reconstruction with the time-domain synthetic aperture focusing technique (T-SAFT) is presented, in which the acoustic velocity is extracted in situ with curve fitting. The operational principle is provided with the help of a numerical simulation, and the confirmation is provided experimentally. In these experiments, an all-optic LUS system was developed by using lasers for both excitation and detection of ultrasound. The acoustic velocity of a specimen was extracted in situ by fitting a hyperbolic curve to its B-scan image. The needle-like objects embedded within a polydimethylsiloxane (PDMS) block and a chicken breast have been successfully reconstructed using the extracted in situ acoustic velocity. Experimental results suggest that knowing the acoustic velocity in the T-SAFT process is important not only in finding the depth location of the target object but also for producing a high resolution image. This study is expected to pave the wave to the development and usage of all-optic LUS for bio-medical imaging.
Funder
Ministry of Trade, Industry and Energy
Ministry of Science and ICT, Ministry of Trade, Industry and Energy, Ministry of Health & Welfare, Ministry of Food and Drug Safety
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry