An In-Field Assessment of the P.ALP Device in Four Different Real Working Conditions: A Performance Evaluation in Particulate Matter Monitoring

Author:

Fanti Giacomo1ORCID,Borghi Francesca2ORCID,Campagnolo Davide1ORCID,Rovelli Sabrina1ORCID,Carminati Alessio1ORCID,Zellino Carolina1,Cattaneo Andrea1ORCID,Cauda Emanuele34,Spinazzè Andrea1ORCID,Cavallo Domenico Maria1

Affiliation:

1. Department of Science and High Technology, University of Insubria, 22100 Como, Italy

2. Department of Medical and Surgical Sciences, University of Bologna, 40126 Bologna, Italy

3. Center for Direct Reading and Sensor Technologies, National Institute for Occupational Safety and Health, Pittsburgh, PA 15236, USA

4. Centers for Disease Control and Prevention, Pittsburgh, PA 15236, USA

Abstract

This study aimed to assess the performance, in terms of precision and accuracy, of a prototype (called “P.ALP”—Ph.D. Air Quality Low-cost Project) developed for monitoring PM2.5 concentration levels. Four prototypes were co-located with reference instrumentation in four different microenvironments simulating real-world and working conditions, namely (i) office, (ii) home, (iii) outdoor, and (iv) occupational environments. The devices were evaluated for a total of 20 monitoring days (approximately 168 h) under a wide range of PM2.5 concentrations. The performances of the prototypes (based on the light-scattering working principle) were tested through different statistical methods. After the data acquisition and data cleaning processes, a linear regression analysis was performed to assess the precision (by comparing all possible pairs of devices) and the accuracy (by comparing the prototypes against the reference instrumentation) of the P.ALP. Moreover, the United States Environmental Protection Agency (US EPA) criteria were applied to assess the possible usage of this instrumentation, and to evaluate the eventual error trends of the P.ALP in the data storage process, Bland–Altman plots were also adopted. The outcomes of this study underlined that the P.ALP performed differently depending on the microenvironment in which it was tested and, consequently, on the PM2.5 concentrations. The device can monitor PM2.5 variations with acceptable results, but the performance cannot be considered satisfactory at extremely low and remarkably high PM2.5 concentrations. Thanks to modular components and open-source software, the tested device has the potential to be customized and adapted to better fit specific study design needs, but it must be implemented with ad hoc calibration factors depending on the application before being used in field.

Publisher

MDPI AG

Reference33 articles.

1. Evaluation of Low-Cost Sensors for PM 2.5 Monitoring: Performance, Reliability, and Implications for Air Quality Assessment;Kurniawati;Res. Sq.,2023

2. Global Review of Recent Source Apportionments for Airborne Particulate Matter;Hopke;Sci. Total Environ.,2020

3. Clements, A., Duvall, R., Greene, D., and Dye, T. (2022). The Enhanced Air Sensor Guidebook, EPA/600/R-22/213.

4. The Global Burden of Disease Due to Outdoor Air Pollution;Cohen;J. Toxicol. Environ. Health A,2005

5. SensEURCity: A Multi-City Air Quality Dataset Collected for 2020/2021 Using Open Low-Cost Sensor Systems;Schneider;Sci. Data,2023

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3