Abstract
Electrical methods are among the primarily studied non-invasive glucose measurement techniques; however, various factors affect the accuracy of the sensors used. Of these, the temperature is a critical factor; hence, the effects of temperature on the electrical properties of blood components are investigated in this study. Furthermore, the changes in the electrical properties of blood according to the glucose level are corrected by considering the effects of temperature on the electrical properties. An impedance sensor is developed and used to measure whole blood impedance in 10 healthy participants at various temperatures and glucose levels. Subsequently, the conductivities of the plasma and cytoplasm were extracted. Changes in the electrical properties of the blood components are then analyzed using linear regression and repeated measures ANOVA. The electrical conductivities of plasma and cytoplasm increased with increasing temperatures (plasma: 0.0397 (slope), 0.7814 (R2), cytoplasm: 0.014 (slope), 0.694 (R2)). At three values of increasing glucose levels (85.4, 158.1, and 271.8 mg/dL), the electrical conductivities of the plasma and cytoplasm decreased. These tendencies are more significant upon temperature corrections (p-values; plasma: 0.001, 0.001, cytoplasm: 0.003, 0.002). The relationships between temperature and electrical conductivity changes can thus be used for temperature corrections in blood glucose measurement.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献