Abstract
We discuss the data sampling frequency, the spectral resolution, and the limit for non-aliasing in the static modulated Fourier transform spectrometer based on a modified Sagnac interferometer. The measurement was performed in a very short 4 ms, which is applicable for real time field operation. The improved spectrometer characteristics were used to investigate the spectral properties of an InGaAs light emitting diode. In addition, The measured spectral peak was shifted from 6420 cm−1 to 6365 cm−1, as the temperature increased from 25 °C to 40 °C, when the operating current is fixed to be 0.55 A. As the applied current increased from 0.30 A to 0.55 A at room temperature, the spectral width was broadened from 316 cm−1 to 384 cm−1. Compared to the conventional Fourier transform spectrometer, the measured spectral width by the static modulated Fourier transform spectrometer showed a deviation less than 10%, and the spectral peak shift according to the temperature rise showed a difference within 2%.