Abstract
Estrogens in personal care products are harmful to customers. Conventional methods such as HPLC and LC-MS require tedious sample pretreatment and long analytical time. Paper-spray ionization mass spectrometry (PSI-MS) is a powerful tool for the determination of compounds with little time and minimal pretreatment procedures. Since most estrogens show poor responses in PSI-MS, we developed a chemical derivatization and PSI-MS method to determinate three estrogens: estradiol, estriol and ethinyloestradiol with estradiol valerate as the internal standard (I.S.). After derivatization with 2-fluoro-1-methyl-pyridinium-p-toluene-sulfonate, the three estrogens could be quantified in seconds. This method showed good linearity in the range of 0.1~30 μg·mL−1, with R2 > 0.999. Their recovery results were all between 85%~115%. The limits of detection (LOD) were 0.04 μg·mL−1, 0.02 μg·mL−1 and 0.02 μg·mL−1 for estradiol, estriol and ethinyloestradiol respectively, which improved around 200, 2000, and 900 times compared to non-derivative PSI-MS. The method could quantitatively determine estrogens in cosmetics.
Funder
NMPA Key Laboratory for Quality Research and Evaluation of Chemical Drugs funding
Subject
Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science