Abstract
A new flow injection spectrophotometric method for the determination of N-acetyl-l-cysteine ethyl ester (NACET) was developed and validated. The method is based on the reduction of Cu(II)-ligand complexes to chromophoric Cu(I)-ligand complexes with the analyte. The studied ligands were neocuproine (NCN), bicinchoninic acid (BCA) and bathocuproine disulfonic acid (BCS). The absorbance of the Cu(I)-ligand complex was measured at 458, 562 and 483 nm for the reactions of NACET with NCN, BCA and BCS, respectively. The method was validated in terms of linear dynamic range, limit of detection and quantitation, accuracy, selectivity, and precision. Experimental conditions were optimized by a univariate method, yielding linear calibration curves in a concentration range from 2.0 × 10−6 mol L−1 to 2.0 × 10−4 mol L−1 using NCN; 2.0 × 10−6 mol L−1 to 1.0 × 10−4 mol L−1 using BCA and 6.0 × 10−7 mol L−1 to 1.2 × 10−4 mol L−1 using BCS. The achieved analytical frequency was 90 h−1 for all three ligands. The method was successfully employed for NACET determination in pharmaceutical preparations, indicating that this FIA method fulfilled all the essential demands for the determination of NACET in quality control laboratories, as it combined low instrument and reagent costs with a high sampling rate.
Funder
Ministry of Science and Education, Republic of Croatia
Subject
Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献