Abstract
This work describes the implementation of the ICP-OES silicon determination method for zinc and lead-bearing materials and waste at the Mining and Metallurgical Works ZGH “Bolesław”. The proposed method was validated. On the basis of linearity tests, it was found that the course of the calibration curve is linear up to a silicon concentration of 100 mg/L, with the determined working range being 0.10–50%. Precision tests, on the basis of which the repeatability was checked, were carried out for nine types of real samples: zinc sulfides, zinc oxides, zinc-lead ore, lead sulfide and zinc-bearing waste. Real samples and six certified reference materials were tested using the ICP-OES radial position. The identified interferences of molybdenum, chromium and vanadium did not statistically significantly affect the measurement results.
Subject
Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science
Reference35 articles.
1. The composition of the continental crust
2. A Background to Silicon and Its Applications;Weiss;AZO Mater.,2001
3. Polish Norm PN-H-04913-05:1975 Chemical Analysis of Zinc Concentrates. Determination of Silica Content (in the Original: PN-H-04913-05:1975 “Analiza Chemiczna Koncentratów Cynku. Oznaczanie Zawartości Krzemionki”)https://sklep.pkn.pl/pn-h-04914-05-1975p.html
4. Performance characteristic of visible spectrophotometer in Analyzing silicon;Adeeyinwo;J. Appl. Sci.,2002
5. The determination of silicate in sea water
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献